NR-ST04 Keyence

Specifications

Model

NR-ST04 *1*2

Type

Strain measurement unit

Input method

Balanced differential input, channels not insulated from each other, input channels insulated from other units (PC)

Number of channels

4 ch *3

Measurement cycle

Maximum sampling cycle: 20 μs (50 kHz) to 60 s

A/D converting system

Successive approximation, simultaneous sampling of all channels

A/D resolution

16 bit

Input frequency band

5 kHz DC max. (−3 dB)

Maximum input rating

±1.0 V

Withstand voltage

Input to system bus/earth

1500 VAC (50/60 Hz) for 1 minute

Input impedance

1 MΩ or more

Input type

Voltage input

±2 mV, ±5 mV, ±10 mV, ±20 mV, ±50 mV

Strain input

±1000 μST, ±2000 μST, ±5000 μST, ±10000 μST, ±20000 μST

Measurable range

±50 mV: −55.00 to +55.00 mV
±20 mV: −22.000 to +22.000 mV
±10 mV: −11.000 to +11.000 mV
±5 mV: −5.500 to +5.500 mV
±2 mV: −2.2000 to +2.2000 mV
±20000 μST: −22000 to +22000 μST
±10000 μST: −11000.0 to +11000.0 μST
±5000 μS: −5500.00 to +5500.00 μST
±2000 μST: −2200.0 to +2200.0 μST
±1000 μST: −1100.00 to +1100.00 μST

Measurement accuracy

Gain accuracy

±0.2% of F.S. (with low-pass filter of 8 Hz and 50 Hz sampling averaged 20 times)

Display resolution

±50 mV: 10 μV
±20 mV: 1 μV
±10 mV: 1 μV
±5 mV: 1 μV
±2 mV: 0.1 μV
±20000 μST: 1 μST
±10000 μST: 0.5 μST
±5000 μS: 0.25 μST
±2000 μST: 0.1 μST
±1000 μST: 0.05 μST

Buffer memory

4M data*4

Warm up time

30 minutes or more

Applicable gauge resistance

Quarter bridge / half bridge system: 120 Ω (A 350 Ω strain gauge requires an external bridge box.);
Full bridge system: 120 Ω to 1 kΩ

Applied voltage

2 VDC (±0.4%)

Balance adjustment

System

Electronic balancing (auto-balancing)

Balance adjustment accuracy

±0.1% of F.S. (with low-pass filter of 8 Hz and 50 Hz sampling averaged 20 times)

Balance adjustment range

±29000 μST (during strain input), ±29 mV (during voltage input)

Gauge ratio

2.0 (fixed)

Low pass filter

Cutoff frequency

Auto, 5 kHz, 3.5 kHz (3 kHz), 2 kHz (1.5 kHz), 1 kHz (750 Hz), 500 Hz (370 Hz), 250 Hz (200 Hz), 120 Hz (100 Hz),
60 Hz (50 Hz), 30 Hz (25 Hz), 15 Hz (12 Hz), 8 Hz (6 Hz), 4 Hz (3 Hz); Value in parentheses is for 20 kHz sampling.

Attenuation characteristics

5th order Bessel filter, −30 dB/oct

Zero-point stability

Temperature characteristics

±2 μST/°C, ±2 μV/°C *5

Chronological changes

±1.5 μST/8 h, ±3 μV/8 h *5

Gain stability

Temperature characteristics

±0.05% of F.S./°C *5

Chronological changes

±0.3% of F.S./8 h *5

Common mode noise reduction ratio

80 dB or more (60 Hz DC max.)

Rating

Power consumption

3.0 W or less

Environmental resistance

Ambient temperature

0 to +40°C 32°F to 104°F

Relative humidity

10 to 85% RH (no condensation)

Weight

Approx. 230 g 8.12 oz (including two OP-51584: NR-ST04 10-pole terminal blocks)

*1 The above specifications are the values when auto-balancing is performed after the warm-up period in an environment with a temperature of 23°C ±3°C 73.4°F ±5.4°F.
*2 ST = strain / μ = 10-6
*3 Max. expansion: 216 ch (with 54 units connected, using NR-X100W), 24 ch (with 6 units connected, using NR-500)
*4 Limited by the NR-X100W buffer memory when using the NR-X100W.
*5 Does not include temperature characteristics and changes over time of built-in bridge resistors. Built-in bridge resistor stability: ±5 ppm/°C or less, ±50 ppm/year

Est. delivery between and    
Estimated delivery dates are estimates only. Int'l customers allow a few more days. For more precise delivery estimates, please contact us at support@plc-direct.com. Contact us in advance for expedited shipping options.
Sale price€783,95
1-year warranty
FREE WORDWIDE SHIPPING
Condition:
€0,00
€0,00
€783,95
Ach Direct Debit American Express Apple Pay Diners Club Discover Google Pay Mastercard PayPal Shop Pay Visa

Specifications

Model

NR-ST04 *1*2

Type

Strain measurement unit

Input method

Balanced differential input, channels not insulated from each other, input channels insulated from other units (PC)

Number of channels

4 ch *3

Measurement cycle

Maximum sampling cycle: 20 μs (50 kHz) to 60 s

A/D converting system

Successive approximation, simultaneous sampling of all channels

A/D resolution

16 bit

Input frequency band

5 kHz DC max. (−3 dB)

Maximum input rating

±1.0 V

Withstand voltage

Input to system bus/earth

1500 VAC (50/60 Hz) for 1 minute

Input impedance

1 MΩ or more

Input type

Voltage input

±2 mV, ±5 mV, ±10 mV, ±20 mV, ±50 mV

Strain input

±1000 μST, ±2000 μST, ±5000 μST, ±10000 μST, ±20000 μST

Measurable range

±50 mV: −55.00 to +55.00 mV
±20 mV: −22.000 to +22.000 mV
±10 mV: −11.000 to +11.000 mV
±5 mV: −5.500 to +5.500 mV
±2 mV: −2.2000 to +2.2000 mV
±20000 μST: −22000 to +22000 μST
±10000 μST: −11000.0 to +11000.0 μST
±5000 μS: −5500.00 to +5500.00 μST
±2000 μST: −2200.0 to +2200.0 μST
±1000 μST: −1100.00 to +1100.00 μST

Measurement accuracy

Gain accuracy

±0.2% of F.S. (with low-pass filter of 8 Hz and 50 Hz sampling averaged 20 times)

Display resolution

±50 mV: 10 μV
±20 mV: 1 μV
±10 mV: 1 μV
±5 mV: 1 μV
±2 mV: 0.1 μV
±20000 μST: 1 μST
±10000 μST: 0.5 μST
±5000 μS: 0.25 μST
±2000 μST: 0.1 μST
±1000 μST: 0.05 μST

Buffer memory

4M data*4

Warm up time

30 minutes or more

Applicable gauge resistance

Quarter bridge / half bridge system: 120 Ω (A 350 Ω strain gauge requires an external bridge box.);
Full bridge system: 120 Ω to 1 kΩ

Applied voltage

2 VDC (±0.4%)

Balance adjustment

System

Electronic balancing (auto-balancing)

Balance adjustment accuracy

±0.1% of F.S. (with low-pass filter of 8 Hz and 50 Hz sampling averaged 20 times)

Balance adjustment range

±29000 μST (during strain input), ±29 mV (during voltage input)

Gauge ratio

2.0 (fixed)

Low pass filter

Cutoff frequency

Auto, 5 kHz, 3.5 kHz (3 kHz), 2 kHz (1.5 kHz), 1 kHz (750 Hz), 500 Hz (370 Hz), 250 Hz (200 Hz), 120 Hz (100 Hz),
60 Hz (50 Hz), 30 Hz (25 Hz), 15 Hz (12 Hz), 8 Hz (6 Hz), 4 Hz (3 Hz); Value in parentheses is for 20 kHz sampling.

Attenuation characteristics

5th order Bessel filter, −30 dB/oct

Zero-point stability

Temperature characteristics

±2 μST/°C, ±2 μV/°C *5

Chronological changes

±1.5 μST/8 h, ±3 μV/8 h *5

Gain stability

Temperature characteristics

±0.05% of F.S./°C *5

Chronological changes

±0.3% of F.S./8 h *5

Common mode noise reduction ratio

80 dB or more (60 Hz DC max.)

Rating

Power consumption

3.0 W or less

Environmental resistance

Ambient temperature

0 to +40°C 32°F to 104°F

Relative humidity

10 to 85% RH (no condensation)

Weight

Approx. 230 g 8.12 oz (including two OP-51584: NR-ST04 10-pole terminal blocks)

*1 The above specifications are the values when auto-balancing is performed after the warm-up period in an environment with a temperature of 23°C ±3°C 73.4°F ±5.4°F.
*2 ST = strain / μ = 10-6
*3 Max. expansion: 216 ch (with 54 units connected, using NR-X100W), 24 ch (with 6 units connected, using NR-500)
*4 Limited by the NR-X100W buffer memory when using the NR-X100W.
*5 Does not include temperature characteristics and changes over time of built-in bridge resistors. Built-in bridge resistor stability: ±5 ppm/°C or less, ±50 ppm/year

Model

NR-ST04 *1*2

Type

Strain measurement unit

Input method

Balanced differential input, channels not insulated from each other, input channels insulated from other units (PC)

Number of channels

4 ch *3

Measurement cycle

Maximum sampling cycle: 20 μs (50 kHz) to 60 s

A/D converting system

Successive approximation, simultaneous sampling of all channels

A/D resolution

16 bit

Input frequency band

5 kHz DC max. (−3 dB)

Maximum input rating

±1.0 V

Withstand voltage

Input to system bus/earth

1500 VAC (50/60 Hz) for 1 minute

Input impedance

1 MΩ or more

Input type

Voltage input

±2 mV, ±5 mV, ±10 mV, ±20 mV, ±50 mV

Strain input

±1000 μST, ±2000 μST, ±5000 μST, ±10000 μST, ±20000 μST

Measurable range

±50 mV: −55.00 to +55.00 mV
±20 mV: −22.000 to +22.000 mV
±10 mV: −11.000 to +11.000 mV
±5 mV: −5.500 to +5.500 mV
±2 mV: −2.2000 to +2.2000 mV
±20000 μST: −22000 to +22000 μST
±10000 μST: −11000.0 to +11000.0 μST
±5000 μS: −5500.00 to +5500.00 μST
±2000 μST: −2200.0 to +2200.0 μST
±1000 μST: −1100.00 to +1100.00 μST

Measurement accuracy

Gain accuracy

±0.2% of F.S. (with low-pass filter of 8 Hz and 50 Hz sampling averaged 20 times)

Display resolution

±50 mV: 10 μV
±20 mV: 1 μV
±10 mV: 1 μV
±5 mV: 1 μV
±2 mV: 0.1 μV
±20000 μST: 1 μST
±10000 μST: 0.5 μST
±5000 μS: 0.25 μST
±2000 μST: 0.1 μST
±1000 μST: 0.05 μST

Buffer memory

4M data*4

Warm up time

30 minutes or more

Applicable gauge resistance

Quarter bridge / half bridge system: 120 Ω (A 350 Ω strain gauge requires an external bridge box.);
Full bridge system: 120 Ω to 1 kΩ

Applied voltage

2 VDC (±0.4%)

Balance adjustment

System

Electronic balancing (auto-balancing)

Balance adjustment accuracy

±0.1% of F.S. (with low-pass filter of 8 Hz and 50 Hz sampling averaged 20 times)

Balance adjustment range

±29000 μST (during strain input), ±29 mV (during voltage input)

Gauge ratio

2.0 (fixed)

Low pass filter

Cutoff frequency

Auto, 5 kHz, 3.5 kHz (3 kHz), 2 kHz (1.5 kHz), 1 kHz (750 Hz), 500 Hz (370 Hz), 250 Hz (200 Hz), 120 Hz (100 Hz),
60 Hz (50 Hz), 30 Hz (25 Hz), 15 Hz (12 Hz), 8 Hz (6 Hz), 4 Hz (3 Hz); Value in parentheses is for 20 kHz sampling.

Attenuation characteristics

5th order Bessel filter, −30 dB/oct

Zero-point stability

Temperature characteristics

±2 μST/°C, ±2 μV/°C *5

Chronological changes

±1.5 μST/8 h, ±3 μV/8 h *5

Gain stability

Temperature characteristics

±0.05% of F.S./°C *5

Chronological changes

±0.3% of F.S./8 h *5

Common mode noise reduction ratio

80 dB or more (60 Hz DC max.)

Rating

Power consumption

3.0 W or less

Environmental resistance

Ambient temperature

0 to +40°C 32°F to 104°F

Relative humidity

10 to 85% RH (no condensation)

Weight

Approx. 230 g 8.12 oz (including two OP-51584: NR-ST04 10-pole terminal blocks)

*1 The above specifications are the values when auto-balancing is performed after the warm-up period in an environment with a temperature of 23°C ±3°C 73.4°F ±5.4°F.
*2 ST = strain / μ = 10-6
*3 Max. expansion: 216 ch (with 54 units connected, using NR-X100W), 24 ch (with 6 units connected, using NR-500)
*4 Limited by the NR-X100W buffer memory when using the NR-X100W.
*5 Does not include temperature characteristics and changes over time of built-in bridge resistors. Built-in bridge resistor stability: ±5 ppm/°C or less, ±50 ppm/year

All PLC Direct products come with a 1-year warranty. The warranty is issued by PLC Direct. Warranty coverage commences at the date of purchase. The warranty is for mechanical defects or failures not caused by human factors including but not limited to (a) incorrect installation, (b) operating the unit in sub-optimal conditions or in scenarios not recommended by the manufacturer, (c) poor unit maintenance, etc. The warranty does not cover programming or software related issues.

If the product is eligible for PLC Direct’s warranty coverage and the customer encounters an issue with the product, we will request the item is shipped to us for diagnosis and examination. Regardless of the cause of product issues, the customer will bear responsibility for the cost of shipping the unit to PLC Direct for evaluation. Where the unit was damaged by human factors, we will offer a repair or replace quote; in cases where damage was caused by human factors, the customer is responsible for the cost of shipping the repaired/replaced unit from PLC Direct to the customer. Where the unit was not damaged by human factors but is covered by PLC Direct’s warranty, we will repair or replace the unit at our cost (or, in rare cases, refund you your purchase); in such cases, PLC Direct will bear the cost of shipping the repaired or replaced product back to the customer. It is up to PLC Direct’s discretion whether to repair, replace, or refund the customer in such cases. Diagnoses can take time to complete based on technician availability and the complexity of diagnosis.

In no event shall PLC Direct be liable for any damages except actual damages up to, but not exceeding, the amount paid to us for the product, including we shall not be liable for any consequential or indirect damages or lost profits whether or not advised of same.

PLC Direct offers free shipping worldwide. Product ship from a various warehouses worldwide based on the customer's and the product's location. The estimated delivery date on the product page is PLC Direct's best estimated assuming a US-based shipping destination.

Regardless of where the product is shipping from, PLC Direct covers all duties for US-based shipping destinations. For non-US-based shipping destinations, customers are responsible for duties as appropriate.

PLC Direct's focus is on balancing price with lead time. Some but not all products can be expedited. If you have questions about delivery ETAs, please contact our team.

Trusted By

"Fast service for obsolete parts"

Increasingly PLC/ HMI control platforms are becoming obsolete and parts are no longer supported, however PLC Direct had the HMI we needed and fast shipping for our system to get back online!

Buyer at GE Health

"Helping us move projects forward"

With supply chain issues increasingly affecting our business, PLC Direct has been there for us to supply the PLC products we need, when we need it. If they don't have it readily available, they find it and provide it quickly and reliably. This has helped us move forward with several projects that would have otherwise taken us many, many months to complete.

Buyer at E-Merge Systems – A Rockwell Integrator